9.5.4.
setting the flow rate range
associated to the 4-20 mA output
The 4-20 mA output provides an electrical current, the value of
which reflects the flow rate measured by the device.
• The signal may be inverted, i.e. the flow rate value associated to the
20-mA current is lower than the one associated to the 4-mA current.
• The current output gives a 22-mA current when the device shows
an operating error.
mA
20
4
20
Fig. 38:
Example of relation between the measuring range and
the current output
0.200
MA
ENTER
ENTER
4ma
*
0......9
To function "FILT"
Enter the flow rate values, in the unit that has been chosen in the
UNIT parameter, related to the 4-20-mA current range.
36
180
l/min
* to move the
1.000
decimal point,
ENTER
simultaneously
20ma
press keys
*
0......9
0......9
English
9.5.5.

choosing the filter of the flow rate

To function "BRGR"
The "Filter" function is used to attenuate the display and 4-20 mA
output, if any, fluctuations when the flow rate varies in the process.
Ten attenuation levels, from 0 (means no attenuation) to 9 (maximum
attenuation of the fluctuations) are available.
Table 4:
Response times of the current output and the display
depending on the filter chosen
Filter N°
0
1
2
and
3 (by default)
4
5
6
Adjustment and functions
Adjustmentandfunctions
2
FILT
ENTER
FILT
0......9
Response time (10% up to 90%)
300 ms
1,5 s
3 s
5 s
7 s
11 s
20 s
ENTER