4. Connect the NIBP connector end to the NIBP port on the Spot-Check Monitor as shown in
figure 2.14.
5. Turn on the Spot-Check Monitor. Press the menu button to go to the settings. Press and hold
the large NIBP measurement button to enter the Pressure Check Mode.
6. Start pumping, and check if the pressure reading on the Spot-Check Monitor matches the
mercury pressure reading.
M3x6 Screw
Figure 2.11
NIBP Port
NIBP Connector Plug
Figure 2.12
Figure 2.14
ENGLISH
Air tube
NIBP Connector
Figure 2.13
Figure 2.15