Télécharger Imprimer la page

Gima PC-300 Manuel De L'utilisateur page 71

Masquer les pouces Voir aussi pour PC-300:

Publicité

Les langues disponibles

Les langues disponibles

71
as
, which means that the adult's blood pressure result is normal. Please re-
fer to the form below for detailed blood pressure levels. Note: blood pressure level indication
is only for "Adult" patient type.
Level
N (Normal)
H-N (High normal)
HT (Hypertension)
NIBP Measurement Principle
The NIBP measurement is based on oscillation technology. The measurement is started by infla-
ting the cuff by a pump automatically after the cuff pressure is high enough to block the arterial
blood flow within the upper arm, then the cuff pressure is deflated slowly, and all the change of
cuff pressure in the deflation process is recorded to calculate blood pressure based on certain
algorithm. The device will judge whether the quality of signal is good enough. If the signal is not
good enough (such as sudden movement or touch of cuff while measurement), the device will
stop deflating or re-inflating, or aborting this measurement and calculation.
Blood pressure range (Unit: mmHg)
SIS <130mmHg, and DIA <85mmHg
130 mm Hg ≤ SIS <140mmHg, and 85mmHg ≤ DIA <90mmHg
SIS ≥140mmHg, or DIA ≥90mmHg
ENGLISH

Publicité

loading