Problem
The device shows
nothing in the
display after you
have fed the test
strip in.
After the test strip
has been fed him,
and the test material
has been applied,
the test procedure
does not start.
Meaning
- The battery is completely discharged.
- The battery is inserted wrongly or
is missing.
- The test strip has been inserted with
the top side facing downwards
or is not completely inserted.
- The device is faulty.
- Insufficient application of blood.
- Defective test strip.
- The test material has been applied after Repeat the test using a new test strip.
the device has switched itself off
(3 minutes after the last user action).
- Defective device.
general information 14
Solution
Replace the battery.
Check that the battery has been
inserted correctly (is the "+" terminal
facing upwards?).
Insert the test strip fully, with the
contacts first and the positive side
facing upwards (page 19).
Please contact our service department.
Repeat the test using a new test strip
and larger drop of blood.
Repeat the test using a new test strip.
You can only apply blood to the test
strip as long as the symbol
in the display.
Please contact our service department.
is flashing
31