Error Messages And Solutions - IBP TD 4222 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

error messages and solutions

The following table explains possible error messages.
Display
Meaning
E-b
Battery empty, no more tests possible.
E-U
Test strip already used, or permanent
electronics error.
E-C
Defective or incorrect test strip,
or test strip inserted wrongly.
E-t
Surrounding, device or test strip
temperatures were too high or too
low in order to carry out a test.
general information 14
Solution
Replace the battery immediately.
Repeat the test using a new
test strip.
If the problem continues, please
contact our service department.
Check to see that the number
shown in the display is the same
as the code number on the test
strip container. Code the device,
or reinsert the test strip.
Repeat the test using a new test
strip once the device and the test
strip have reached the operating
temperature (+ 10 °C to + 40 °C).
29