how to check using control solutions

1. Feed your test strip into the test strip opening with the contacts first, and
the device display facing you (see Page 19). You should ensure that the
test strip is firmly located in the device to achieve exact measured results.
The device switches itself on automatically. The display will show "CHK"
and the "Test strip" symbol. After this, the surrounding temperature, the
symbol
and the code number are displayed. You should calibrate your
device (see Page 10) if the code number shown in the display is not the
same as the code number on the test strip container.
2. If you press the Menu button once, "CTL" is shown in the display. If the
"CTL" symbol is being shown in the display, any blood sugar levels
measured will not be saved to the internal memory. This means that you
will not be influencing your measured values statistics when carrying out
tests with the solutions. If you want to save the following measurement to
the memory, however, press the Menu button again and "CTL" will no
longer be shown in the display.
Warning: Every time you carry out a test with the control solution, you should always activate the
CTL mode beforehand so that your saved measured value statistics are not falsified!