3. Shake the control solution well before using it. Unscrew the cap and
squeeze one drop out. Wipe the first drop away, and then squeeze another
one out. Put the drop onto your fingertip (Fig. 7).
Note: In order to make sure that the control solution is not contaminated by
contact with the test strip, do not apply the drop directly to the strip, but
first apply it to something clean or to your fingertip.
4. Apply the drop onto the absorption gap. The solution is sucked into the
gap. When the control window is completely full of the solution, the blood
sugar measuring device starts a 10 second countdown.
5. The glucose measurement takes place after the countdown has finished,
and then the result is shown in the display. Check to see that the result is
the same as the stipulated control solution value. This value is printed on
the test strip container.
Warning: If you achieve incorrect measurement results with the control solution which deviate
from the stipulated value time after time, the measurement system is no longer working
correctly. In this case, you should stop using the device to measure your blood sugar
level. Please see the chapter entitled "troubleshooting" on Pages 29-31. If you are still
unable to solve the problem, please contact our customer service department (see
guarantee page).