12.2. Calibrating the ORP probe
1) Open the "ORP calibration solution" vial.
2) Turn off the filter (and therefore the electronics unit).
3) If the probe is already installed :
a) Remove the probe from the probe connector, without disconnecting it.
b) Remove the probe holder nut and replace it with the stopper provided.
If the probe is not already installed :
Connect the probe to the electronics unit.
4) Turn on the electronics unit.
5) Go to the "
Electrolysis ORP Calibration
6) Navigate through the menus following the below instructions :
Electrolysis
ORP Calibration
N
ORP Calibration
Solution
470 mV
N
ORP Calibration
In Progress
(Wait a few seconds)
ORP Calibration
Success
→ If calibration fails, the screen will display the below message :
ORP Calibration
Failed
If this happens, navigate through the menus a second time following the above instructions.
If calibration fails a second time, replace the probe and carry out another calibration.
" menu (see chapter 10).
→ Insert the probe into the ORP calibration solution, then wait 5 minutes.
→ Do not touch the probe.
→ Calibration is finished.
a) Rinse the probe under running water, then let the water drip off without
drying with a cloth.
b) Install the probe into the probe connector.
27