portable and mobile RF communications equipment and the Elexo XXL
The Elexo XXL is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The cus-
tomer or the user of the Elexo XXL can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Elexo XXL as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watt (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
88 | Benutzeranleitung | Elexo XXL Dusch- und Toilettenstuhl
Recommended separation distance between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d =1,2 √P
0,12
0,38
1,2
3,8
12
m
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d =1,2 √P
d =2,4 √P
0,12
0,24
0,38
0,76
1,2
2,4
3,8
7,6
12
24