Guidance and Manufacturer's declaration-electromagnetic immunity
The Elexo XXL is intended for use in the electromagnetic environment specified below. The customer or the user of the Elexo
XXL should assure that it is used in the following environment:
Immunity Test
IEC 60601-1-2 Test
Level
RF Common mode/
3 Vrms
Conducted Susceptibility
150 kHz to 80 MHz
IEC 61000-4-6
Radiated RF Electromag-
netic Field
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To as-
sess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If
the measured field strength in the location in which the Elexo XXL is used exceeds the applicable RF compliance level above,
the Elexo XXL should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the Elexo XXL.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Compliance Level
3 V
3 V/m
Elexo XXL Shower-toilet chair | Instructions for use | 43
Electromagnetic Environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Elexo
XXL, including cables than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance.
d =1,2 √P
d =1,2 √P
80 MHz to 800 MHz
d =2,4 √P
800 MHz to 2,5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol: