Log; 6.3.10 Info; Automatic Logout - Sartorius Biowelder Total Containment Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Operation
32
Operating Manual Biowelder
6.3.9

Log

For general information about logs, please read Chapter 9.7 Log
Through the Info menu logs can be downloaded to the computer.
Selection
Weld (CSV)
Weld (HTML)
All
When outputting logs via Web browser, not all data saved in the device is
outputted . The most current log is output up to a file size of around 500 kB .

6.3.10 Info

Detailed information can be downloaded in this menu. The following table
lists this information:
Function
Meaning
State
The current status of the device can be downloaded
and viewed via the Download button.
Download
A log file can be downloaded and opened by pressing
log files
the corresponding selection button. Further details can
be found in Chapter 6.3.9 Log.
Maintenance
Informs you about the abrasion condition of
maintaining parts. If the wearing down of one or more
parts has reached ≥100%, an information will occur on
the operator screen. Refer to Chapter 5.5 Maintenance.
Date and
see Chapter 6.3.7 Date and time.
Time
6.4

Automatic Logout

Login and automatic login is already described in chapter 6.2.9 User
management and 6.3.8 User management.
The automatic logout feature is based on CFR21 Part 11. After a certain time
without user interaction, the machine logs out automatically. Changes not
saved will be discarded. However, there are certain exceptions.
No user interaction is required during a process (e.g. welding). The time
starts counting after the weld is completed. Weld log screen or screen for
removing the blade do not count to the process. If the timer runs out during
one of those screens, the login screen will appear after the last screen
(remove blade).
The same situation is by the validation or initialisation. No user interaction is
needed as long as the BWTC is doing some work (moving motors or
heating). After the work
is done (waiting for the user to do something) the logout timer is running.
The login screen will appear immediately after time run out or after some
ending up user interaction (e.g. removing the blade).
The timer can be adjusted in the web interface under general parameters. If
a user has the automatic login, the automatic logout is deactivated.
Total Containment
®