Informational Icons - Inogen Rove 6 Manuel De L'utilisateur

Table des Matières
Les langues disponibles

Les langues disponibles

7.4 INFORMATIONAL ICONS

The following displayed icons are not accompanied by any audible feedback or any visual change in the
indicator lights.
Display Icons
Description & Action (if needed)
Flow setting: "X" represents the selected flow setting (e.g., setting 2).
Please wait indicator: This symbol will appear while the concentrator starts up.
Following a brief start-up sequence, a warmup period up to 2 minutes will initiate.
During this time-period the oxygen concentration is building to but may not have
reached specification.
Time remaining on battery charge: "HH:MM" represents the approximate time
remaining on the battery charge in hours:minutes (e.g., 1:45).
Battery charge and charging status: This symbol indicates that the battery is installed
and is charging. For a complete list of battery charging symbols, see 'charging the
battery with the concentrator' (section 6.8).
Battery level status: This symbol indicates the battery level (about 50% in this
example). Refer to 'checking the battery status when installed on the device'
(section 6.6).
Battery % charged: This symbol will be displayed when the concentrator is plugged
in and is being used to charge a battery (not being used for oxygen production). It
is normal to see a fully charged battery read between 95% and 100% when external
power is removed. This feature maximizes the useful life of the battery.
Sieve (columns) reset: This symbol is displayed when column maintenance is
required and once the replacement columns have been installed.
Sieve reset success: This symbol is displayed once the sieve columns have been
successfully reset.
Data log transfer in progress or update in progress (app only): This icon is displayed
during all data log transfers and software updates initiated through the Inogen
Connect App.
Data log transfer success (app only): This icon is displayed after data log transfers
have been successfully completed through the Inogen Connect App.
The following displayed icons are accompanied by a single, short beep.
Please wait, shutting down: Power button has been pressed for 2 seconds.
Concentrator is performing system shut down.
Life Clock (HH:MM), software version & serial number display (Vx.x:SN):
The Life Clock, software version & serial number will be displayed when the
'No-breath-detect' audible alarm button (bell button) has been pressed for five
seconds while the concentrator is running.
©2024 Inogen, Inc. All rights reserved.
Page 20 of 120
96-13115-50-01 B
Table des Matières
loading

Table des Matières