INDICATIONS FOR USE
• Maxima Turbo complies with IEC 60601-1-2:2014 (Edition 4.0)
• Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify that they are operating normally. Max power cord length is 160 cm.
• Use of accessories and cables other than those specifi ed or provided by Maxima Turbo could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
• Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the Maxima Turbo S, including cables specifi ed by Maxima Turbo. Otherwise, degradation of the performance of this
equipment could result.
• The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
• The user is responsible for the installation, correct operation and maintenance of the device in accordance with instructions provided in this user
manual. If needed, contact the service or the supplier of the device.
• The sterilizer is not intended for sterilizing liquids, biomedical waste or pharmaceutical products.
• The sterilizer must not be used if explosive gases or vapours are present in the air.
• After the cycle is completed, the load is hot. Remove tools or packs from the chamber using appropriate thermal gloves or equipment that prevents
burns.
• Do not remove the rating plate or any other elements of labeling from the device.
• Follow guidelines for preparing tools for sterilization.
• Pouring water or other liquids on the device may cause a short-circuit.
• Prior to inspection, maintenance or servicing, turn off the device and disconnect it from the power source.
• Servicing may only be performed by trained service personnel and using original replacement parts.
• The device is intended for indoor use only.
• Max working high elevation of the device is 2,000 m above sea level.
• Pollution Degree 2: Normally only nonconductive pollution occurs. Temporary conductivity caused by condensation is to be expected.
• Overvoltage Category II.
• If the equipment is used in a manner not specifi ed by the manufacturer, the protection provided by the equipment may be impaired.
• Use only detachable cord that rated equal or greater to the equipment electrical ratings.
Read this Operating Manual carefully before using this device. Install and operate the device strictly as specifi ed herein. Comply with all safety
requirements for the device. This will ensure proper and safe operation of this device. Any other application, inconsistent with this manual, may lead
to dangerous accidents. Restrict unauthorised personnel access to the device and train the personnel handling the device. An operator of this device
is any person who, by training, experience and knowledge of applicable reference standards, manuals and occupational health and safety regulations
has been authorised for the essential operation with the device and who is capable of identifying and avoiding the hazards related to operation of this
product.
Always append this Operating Manual with the device if transferred to a new owner. The Operating Manual contains detailed information about
assembly, installation, initial start-up, use, repairs and maintenance of the device. If the device is used as intended, this Manual will provide suffi cient
guidance to qualifi ed personnel. Keep this Operating Manual close to the device and easily accessible at all times. As required by continuous improve-
ment of the product, the manufacturer has the right to amend this Manual or make changes to the device without prior notice. Henry Schein shall not
be liable for damage incurred during the wait for warranty service, any damage to the Customer's property other than this device, or errors caused by
improper installation and/or improper operation of the device.
Detailed recommendations, coutnerindications and warnings are
described in the relevant sections.
PRODUCT STRUCTURE AND ACCESSORIES
22 in (561 mm)
9.9 in (252 mm)
Maxima Turbo - 3