Decorative illumi-
nation
Decorative illumination
color change
Counters
The Maxima Turbo sterilizer records the number of processes performed. This lets you know when you need to replace consumable parts.
Counter section with the counters reset to zero. The number of processes performed is on the left, while on the right is the number at which the com-
ponent should be replaced or a service inspection should be performed. After replacing the fi lter, the user can reset the values by pressing the RESET
button.
During the fi rst process (Vacuum, Helix, FAST, 134/273, 121/250), the unit will record the current date as the date of the fi rst process (bottom line on
the counter screen). The device will count processes and days since the fi rst startup.
Displaying the counter values in yellow or red does not block the operation of the unit. However, exceeding the required replacement time may have a
signifi cant impact on the operation of the unit and the sterility of the load. For replacement of individual components, please contact the manufacturer
or supplier.
No.
1
Restart
The restart of the process is forced when the user stops the process by pressing the STOP fi eld, in the case of a power outage or an error during the
process, for example lack of supply water.
If the STOP fi eld is selected, the following messages are displayed alternately:
• The user has stopped the process.
• Equalizing the pressure in the working chamber.
• Process is incorrect, which means that the load is not sterile.
When the pressure in the working chamber is equalized, the following messages appear alternately on the display. Now you can open the device
freely. After opening the chamber, the screen appears.
Illumination during a
process step
Turning the LED illumina-
tion on and off
Name
HEPA fi lter and Door with
gasket
Replacement frequency
(cycles)
1000
Decorative illumination color
change
Yellow
Early replacement call
(cycles)
980
Red
Latest replacement after
(cycles)
1000
Maxima Turbo - 17