"Overpressure during standby"
TECHNICAL DATA
Technical data
Power supply
Installed power
Maximum electric current consumption
Operating pressure
Maximum pressure
Maximum process temperature
Process chamber capacity
Mass
Process chamber dimensions (LxWxH)
External device dimensions (LxWxH)
Protection rating
Noise level
Process data archiving
SERVICE
For items that require service, including warranty service, please send to Henry Schein® Handpiece & Small Equipment Solutions.
Henry Schein Handpiece & Small Equipment Solutions
620 S. Placentia Ave.
Placentia, CA 92870 USA
Phone # 800-235-6337 Prompt 1
WARRANTY
This product is designed for use in a dental office and this warranty is not applicable to other uses. This product is warranted against manufacturing
defects in materials and workmanship for 2 years from date of purchase.
Henry Schein® will repair or replace product at its sole discretion. This warranty shall be limited to replacement or repair of the product or its parts
and shall not extend to any other claims, including but not limited to loss of profit, cost of removal or replacement, incidental or consequential dam-
ages or other similar claims arising from use of this product. Damage to product resulting from acts of God, faulty installation, misuse, tampering,
accident, abuse, negligence, alterations, unauthorized repairs or problems unrelated to materials and workmanship are not covered by this warranty.
Hypertension in standby mode
Maxima Turbo
110-120 V/60Hz
1.6 kW max.
15 A
2.1 bar / 30.5 psi
2.45 bar max / 35.53 psi
137°C (278°F)
2.7 l / 0.7 gal
15 kg / 33 lb
292 x 192 x 45 mm / 11.4 x 7.5 x 1.8 in
561 x 252 x 162 mm / 22 x 9.8 x 6.3 in
IP20
49dB(A)
USB drive
The reason for this error is that the hot sterilizer
is left with the chamber closed (e.g. overnight).
As the sterilizer cools down a vacuum is creat-
ed in the chamber which causes a startup error
Wait until the device has equalized the pressure
automatically - the message will disappear
automatically