tion result.
Error codes
The table below contains error codes that may appear when working with the sterilizer
No
Error code
1
"Chamber over temperature"
2
"Steam gen. Over temperature"
3
"Process over temperature"
4
"Overpressure error"
5
"Sterilization pressure too low"
6
"Sterilization temp. Too low"
7
"Too high pressure during drying"
8
"Too many steam pulses / no water
9
"Drainage error"
10
"Chamber heating error"
Description
Maximum chamber temperature exceeded
Steam generator temperature too high
Process temperature too high
Pressure error
Drying pressure too low
Sterilization temperature too low
Drying pressure too high
Too many steam pulses. No feed water.
Clogged drain
Chamber heating error
Recommendations
Contact with the service
• Too high weight of sterilized instruments -
repeat the process with less instruments (max.
0.5 kg)
• Contact with the service
Contact with the service
Contact with the service
• Check that there is water in the bottle with the
blue hose
• Correct the position of the blue hose so that the
end is completely submerged in water. Add a
sinker to eliminate the problem in the future
• Check that the water supply hose (blue) is not
damaged (After correcting the position / replac-
ing the hose or refilling with water, restart the
machine)
• Contact with the service
• Check the water level in the bottle with the blue
hose
• Check that the red tube is not pointing up along
its entire length, creating the so- called air trap
• Contact with the service
• Make sure the red drain hose is not immersed
in water. The hose must not be kinked, the
liquid must flow down by gravity
• Check that the weight of the sterilized instru-
ments is not too high
• Contact with the service
• Check the water connection to the "water in"
connector
• Check the distilled water level in the feed water
tank (blue plug)
• Check that the weight of the load does not
exceed the allowable weight
• Contact with the service
• Clogged drain Check the level of the wastewa-
ter and the connection of the hoses
• Check the level of the used water in the bottle
with the red cap. If the bottle is full, discard the
used water
• Check that the red hose is not kinked and that
it is pointing downwards along its entire length
• Check that there is no debris in the outlet open-
ing (inside the chamber)
• Contact with the service
• Mains voltage too low - consult an electrician
for the location where the autoclave is to be
installed
• Contact with the service
Maxima Turbo - 21