Program selection
Depending on the type of load to be sterilized, the user must select the appropriate program - in accordance with the instructions of the sterilizer
manufacturer and the load manufacturer's recommendations.
Maxima Turbo program 134°C (273°F) FAST
Type of load
Unwrapped instruments only
Process temperature
134°C (273°F)
Pre-vacuum number
1
Sterilization duration
4:00 min
Drying duration
-
Total process duration*
100g: 7 min
Class
S
Load Description
Solid Objects, non-porous objects,
simple instruments (such as scissors,
handles pliers, chisels, probes, etc.)
and dental handpieces; unwrapped
After switching on the device, the start screen appears on the display. To go to the next screen, press the screen once with your fi nger (anywhere).
On the next screen, we can go to sterilization, test the device or open the information menu. The following sterilization programs can be selected from
the main menu: 121 ° C (250°F) , 134 ° C (273°F) , and 134 ° C (273°F) FAST.
Test programs
134°C (273°F)
Wrapped or unwrapped instruments
134°C (273°F)
3
4:00 min
3 min
100g: 13 min
B
Solid objects, small porous objects,
simple objects recessed, nar-
row-clearance items, dental hand-
pieces, and textiles; wrapped and
unwrapped
After switching on the device, the start screen appears on the display. To go to the next
screen, press the screen once with your fi nger (anywhere).
Sterilization programs
Selection of sterilization programs
121°C (250°F)
Wrapped or unwrapped instruments
121°C (250°F)
3
30 min
5:30 min
100g: 45 min
B
Solid objects, small porous objects,
simple objects recessed, nar-
row-clearance items, dental handpiec-
es, textiles, and plastics; wrapped and
unwrapped
Settings Menus
Maxima Turbo - 11