By selecting the fi eld, we can return to the welcome screen. In the event of an error, we must additionally enter the 4-digit security code 0000. Enter-
ing this code is tantamount to the operator's declaration that he is aware that the sterilization process has not been carried out properly and that the
batch is non-sterile.
DEVICE MAINTENANCE
In order to ensure the correct operation of the device, the user is obliged to perform the following maintenance actions.
Cleaning the tray
Keeping the tray clean helps to ensure proper operation of the appliance. The tray and its technical condition are a good indicator of using the correct
water. A stony, brown tray indicates the use of poor quality water. It is recommended to clean the inside of the tray on a weekly basis with a mild
detergent that does not add chlorine and does not react with aluminum. After cleaning, the tray requires thorough rinsing with water. In order to clean
the tray properly, it should be removed from the front of the device.
Cleaning the process chamber
Keeping the chamber clean helps to ensure proper operation of the device. It is recommended to clean the inside of the process chamber once a week
with a mild detergent without the addition of chlorine. After cleaning, the chamber should be wiped dry with a soft cloth. Cleaning should be per-
formed on a cold chamber.
Cleaning of external surfaces
Cleaning the external parts of the device should be done with a soft cloth moistened with water and a mild detergent (no chlorine added and not
reacting with plastics, varnish coatings, aluminum). Strong detergents should not be used. The use of mild detergents to maintain the device does not
affect the risk of toxic components coming into contact with device components.
Cleaning the gasket
It is recommended to clean the gasket each time after 100 processes. Use warm, clean water and microfi ber to clean the gasket (microfi ber with sil-
ver particles is allowed). Blunt and sharp cleaning tools are not allowed. Cleaning with chemicals is not allowed. Carry out cleaning on a cooled down
device, after opening the drawer. Be careful not to bend the drawer. After cleaning, leave the device open until the gasket is dry. During this time,
protect the device against damage.
Replacement of consumables
The sterilizer is equipped with a high performance sterilization system. A message on the screen informs the user when each component should be
replaced. If the sterilizer is in regular operation, alternating replacement screens will appear after the welcome screen has been pressed. The replace-
18 - Maxima Turbo
Lift the tray up gently and pull away from the front face. The mounting pins have notches
into which the drawer fi ts. Before reinstalling the tray into the unit, drain the tray and slide
it over the front face pins and press down gently to lock.