Circuit For Differential Pressure Measurement - HBM P3MB Notice De Montage

Capteur de pression absolue
Masquer les pouces Voir aussi pour P3MB:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

P3.../P3 Top Class
6

Circuit for differential pressure measurement

Two transducers with the same full scale value can be readily interconnected,
in order to determine the relevant differential pressure signals. To do this, the
excitation voltage connections are connected in parallel and the measuring
voltage connections crossed over. The resultant resistance is now 175 Ω.
P
U
Transducer 1
P
U
Transducer 2
Fig. 6.1: Circuit diagram for differential pressure measurement
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Transducer 1
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Transducer 2
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ
The available differential pressure signal is: U
A2239-3.0 en/de/fr
bk
bu
gn
gy
rd
wh
bk
bu
gn
gy
rd
wh
Excitation voltage
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
bk
bu
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
with
with
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
bk
bu
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
VKK terminal box
Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Measuring voltage
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ Ñ
wh
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
with
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
rd
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
.
= 1/2
(U
A
to the amplifier
bk
bu
gn
U
U
A
gy
B
wh
rd
U
= differential
A
pressure signal
U
= bridge excitation
B
voltage
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
rd
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
with
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
wh
Ñ Ñ Ñ Ñ Ñ Ñ Ñ
- U
)
A1
A2
19
HBM

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

P3mbpP3 top classic

Table des Matières