The alarms which occured are indicated on a list.The following information is separately made for disposal for each error type:
Error type
Error counter
Date / time /
alarm start
Date / time /
alarm end
An arrow marks the respective active list entry.
In pressing the keys [+/-] the arrow can be moved.
In pressing the [ENTER] key the respectively active list entry is cancelled. Afterwards the alarm information for
this kind of error is not available any more.
If no alarm occured or if all list entries were cancelled [„....<ESC>...."] appears on the display.
The diagnosis menu can only be quit with [ESC].
F3
In a list the alarm parameters for temperature, speed, door contact and voltage loss can be adjusted.
An arrow marks the respectively active parameter.
In pressing the keys [+/-] the arrow can be displaced.
In pressing the [ENTER] key the YES / NO definitions are directly changed or the numeric values can be released for modification
(value flashes).
In pressing the keys [+/-] the respectively flashing value can be set.
In pressing the [ENTER] key the adjusted value is confirmed.
In pressing the [ESC]-key you jump back to the SETUP1-menu.
26
DIAG:
e. g. speed or temperature alarm
Number of the alarms which occured since the last acknowledgment
Date and time of the alarm start (only of the alarms which occured last)
Date and time of the alarm end (only of the alarms which occured last)
ALERTS:
Diagnosis of the alarm messages which occured
Alarm definition