F3
After the pressing of [F3] the "TIMER" [TIME] can be adjusted.
In pressing the keys [+/-] the desired operating time in [h : min] is adjusted
Confirm input with [ENTER].
The operating time can be set between 00h:01min and 99h:59min as well as "CONT." (continuous operation).
F4
After pressing of [F4] the cooling is switched on / off.
Switching the cooling on is only possible at temperature set values < 38.5°C!
If a higher temperature value is entered and confirmed by pressing [ENTER] while the refrigerator is on,
a brief signal sounds and the refrigerator will be switched off.
MODE
The respective "active" menu is indicated, here: "MAN" for manual menu.
12
TIME:
Set-up of the TIMER
COOLING:
Switching the cooling on and off (only at "UHK version ")
MODE:
Display field