4. Safety functions

4.1. Emergency stop button

(only relates to electrical adjustable versions)
In the centre of the undercarriage, there is an emergency stop
button . Operate this button in an emergency; if the upward
or downward movement must be stopped immediately . After
the emergency stop has been pressed, all electrical operation
will cease to function . Operation may be continued once the
situation is stable again . Turn the emergency stop a quarter
turn to unlock it again .
First check to ensure the hazard has been removed
or the problem resolved, before resetting the emergency stop.

4.2. Emergency lowering button

(only relates to electrical adjustable versions)
If the controls stop working and the shower stretcher /
dressing table is in a raised position, it can be lowered using
the emergency lowering button . This button is located to the
left of the emergency stop button (See figure 12) .
First check to ensure that the emergency button has not been
pressed!
- Remove the plug from the manual control .
- Press the emergency lower button (to the left of the emergency stop) .
The shower stretcher / dressing table will now return to the lowest position (make sure
no persons or objects are under the stretcher) .
Take the less able person from the shower stretcher / dressing table and make sure the
device is not used until it has been repaired . Place the shower stretcher / dressing table
in a safe place and indicate that it is out of order .
Contact the technical service or Lopital's service department .
E-mail: service@lopital .nl
44 | Instructions for use | Shower stretcher / Dressing table