Cleaning Controls; Lighting Maintenance; Electrical Connections; How To Connect The Cables To The Batteries - Quickie Salsa M2 Mini Manuel D'utilisation

Masquer les pouces Voir aussi pour Salsa M2 Mini:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

13.8 Cleaning controls

Clean the control system and the joystick with a cloth
dampened with diluted detergent.
CAUTION!
Be careful when cleaning the joystick.
Never use abrasive or spirit based cleaners
Speciality controls
Ensure that wafer boards, joysticks (all variants), head
arrays and switches (all variants) are cleaned with a
mild disinfectant and a cleaning cloth to avoid any cross
infection possibilities. Following removal from wheelchair
regularly wash the sip & puff mouth piece and tube to
maintain cleanliness and functionality.
WARNING!
Ensure control is switched off before cleaning.

13.9 Lighting maintenance

WARNING!
Maintaining lights and indicators is a safety critical task.
If your lights and indicator system shows a defect, please
contact your Sunrise Medical Approved Dealer.
All lights and indicators are state of the art LED low
energy, no maintenance units. There are no light bulbs
used. The high intrinsic reliability of these units means
that they are extremely unlikely to fail under normal
usage. If a failure should occur, (may be as a result of an
impact), the damaged light assembly must be replaced in
its entirety. The individual LED's cannot be replaced.
Rev.3.0
CAUTION!
We recommend the use of Sunrise Medical authorised
spare parts only.
NOTE: Be advised that all lighting circuits are
electronically protect ed. In the event of a short circuit
current will be limited to a safe level. Once the fault is
removed the system resets automatically.

13.10 Electrical connections

When inspecting electrical connections, pay attention to
the battery connections, the connection of batteries to
power loom and plug in sockets for the joystick, control
box and lights and indicators.

13.11 How to connect the cables to the batteries

Also refer to section 11
WARNINGS!
If you are in any doubt, please contact your Sunrise
Medical authorised dealer.
Batteries are heavy, please use correct lifting
techniques and always use the straps provided to lift
the batteries.
No smoking.
Remove any conductive jewellery such as neck
chains, bracelets, watches etc.
Before connecting the battery terminals it is very
important to identify the correct connection for each
battery terminal. Failure to adhere to this will result in
the wheelchair not functioning or a ruptured fuse.
Connect all battery terminals before inserting the
connectors.
After connecting battery terminals and connectors
ensure all wiring is tidy and cannot get entangled in
any moving parts.
67
SALSA M
Mini
2

Publicité

Table des Matières
loading

Table des Matières