Indications For Use
This air-powered high speed handpiece is intended to be used for general dentistry procedures.
Contraindications
None Known.
Warnings
S S T T E E R R I I L L I I Z Z E E A A N N D D L L U U B B R R I I C C A A T T E E B B E E F F O O R R E E U U S S E E , , S S U U P P P P L L I I E E D D N N O O N N - - S S T T E E R R I I L L E E . .
WARNING: This product is intended for use by dental professionals only. Before operating
handpiece, carefully read and follow these instructions and save them for future reference.
Observe all cautions and warnings, including:
• Handpieces should be heat-sterilized between patients to prevent cross-contamination which
could result in serious illness or even death from infectious organisms such as HIV (the virus
that causes AIDS) or hepatitis B.
• It is the responsibility of the Dental Healthcare Professional to determine the appropriate use
of this product and to understand the health of each patient, the dental procedures being
undertaken and industry and governmental agency regulations, recommendations and
requirements for safe practice of dentistry.
Precautions
• Do not cut with the handpiece unless you are sure the bur is secure. Do not extend the bur
outward from the handpiece head. A loose or extended bur could eject from the head or break
and cause personal injury.
• Do not use this handpiece within the oral cavity when a soft tissue wound is present.
Air pressure may force septic materials into these tissues.
• Midwest turbines should not be rebuilt. Replacing bearings only could result in unexpected
failure and patient injury due to different wear factors of new vs. old components.
Adverse Reactions
None Known.
Installation
Checking Handpiece Hose
Make sure all tubings supplying the handpiece hose are properly connected within the dental unit.
Be sure the air and water are delivered to the handpiece as shown in the illustration below.
5-Hole Fiber Optic
Fiber Optic
or Bulb
Air Requirements
The handpiece delivers designed speed and torque when operated with clean, filtered, moisture-free air at the
recommended operating pressure of 30 p.s.i. (2.0 Atm.). Air pressure should be measured at the handpiece
end of the hose.
Water Supply Recommendations
It is highly recommended that all dental water supply systems conform to applicable CDC (Centers for Disease
Control and Prevention) and ADA (American Dental Association) standards, and that all recommendations be
followed in terms of flushing, chemical flushing and general infection control procedures.
Initial Maintenance
Follow the Maintenance procedure, below, before using your new Tradition or Quiet-air handpiece.
Maintenance
Important: Only use Midwest
®
products recommended by DENTSPLY to routinely maintain your
handpiece. See Parts section for ordering information.
There are two Maintenance procedures for your Tradition or Quiet-air handpiece: (1) for maintenance using the
Midwest Plus
Maintenance Kit (REF 380145) ; (2) for maintenance using the Midwest Plus
380080). Follow the appropriate maintenance procedure, at right, before using your new Tradition or Quiet-air
handpiece.
These procedures recommend the steps to follow between patient use to prevent cross-contamination and to ensure
continued trouble-free operation of the handpiece:
1
Maintenance Using
Midwest Plus
Maintenance Kit
Clean
1. Remove bur.
2. Spray Midwest Plus™ Handpiece Cleaner liberally:
(a) into chuck
(c) onto exterior surface of handpiece (including fiber
optic bundles on both ends of the handpiece).
Quiet-air
®
4-Hole Non-Fiber Optic
Aerosol Spray (REF
(b) into drive air tube
gold tube or
shorter of the 2
larger tubes
Midwest Tradition
High Speed Handpiece
Fiber Optic / Non-Fiber Optic
(Tradition
REF 750045 / 750044) (Quiet-air
®
(Tradition
L
®
REF 780045 / 780044) (Quiet-air
(Tradition
PB
REF 790045 / 790044)
®
Installation, Maintenance and Operation Instructions
Scrub and Rinse
Scrub and rinse under warm tap water. Make sure fiber optic bundles are scrubbed well.
Expel Cleaner and Dry
Run handpiece on Midwest Handpiece Air Station (REF 380800) or other air source for 30 seconds or until
handpiece reaches normal speed, and all excess cleaner is expelled. Handpieces with bur tool chucking system
(Tradition REF 750045/750044 or Quiet-air REF 730014/464004) should always be covered with a Spray Guard
or have a bur installed before running. Dry thoroughly.
Lubricate and Expel
1. Lubricate by placing 3 drops of Midwest Plus Handpiece Lubricant (REF 380130) into drive air tube (shorter
of two larger tubes).
2. Run handpiece for 30 seconds or until it reaches normal speed, and until all excess lubricant is expelled.
Wipe off any excess lubricant on outside of handpiece with a dry gauze pad or tissue.
2
Maintenance Using
Midwest Plus
Aerosol Spray
Scrub and Rinse
Remove bur. Scrub and rinse under warm tap water. Make sure fiber optic bundles are scrubbed well.
Clean, Lubricate and Expel
Clean and lubricate with one quick spray of Midwest Plus™ Aerosol Spray (REF 380080) using the Midwest
Backend Attachment nozzle (REF 380086) into the drive air tube of the handpiece. The drive air tube is the
shorter of the two larger tubes and is gold on many handpieces. Run the handpiece on Midwest Handpiece Air
Station or other air source for 30 seconds or until normal speed is reached, and all excess spray is expelled.
This handpiece with a bur type chucking system should always be covered with a Midwest Spray Guard or have
a bur installed before running. Dry thoroughly.
Bag • Autoclave
Insert handpiece into sterilization bag and load into autoclave. (DENTSPLY recommends using sterilization bags
with at least one paper side.) Handpieces should be processed in an autoclave set for a minimum of 15
minutes at sterilization temperature (132ºC/270ºF).
For cassette chamber saturated steam devices, handpieces should be processed unbagged for a
minimum of 3.5 minutes at sterilization temperature (135ºC/275ºF).
Keep handpiece in bag until ready to use in the operatory. No relubrication is necessary.
Other Maintenance
Ultrasonically Clean Handpiece Head
As needed, only if sluggishness occurs:
1. Place handpiece head down in beaker of Midwest Plus Handpiece Cleaner.
2. Ultrasonically clean for 10 minutes.
3. Continue from Scrub and Rinse, above.
4. In the lubricating step, also place 1 drop of Midwest Plus Handpiece Lubricant (REF 380130) in chuck
(for Push Button and Latch models: REF 790045 / 790044 / 780045 / 780044 / 780014 / 484004).
Cleaning Water Spray Tube
Run the cleaning wire back and forth inside the inner water spray tube and around outer chip air tube to loosen
and remove mineral build-up at spray tip.
Operation
(Standard: REF 750045 / 750044 / 730014 / 464004)
Never operate the handpiece under the following conditions:
1. In the presence of a soft tissue wound.
2. With a bur extended out from handpiece head.
3. Under "no load" conditions for extended periods of time.
4. At air pressure exceeding 30 p.s.i. (2.0 Atm.).
To operate the handpiece: Simply depress the foot control on your dental unit. For water spray and chip air,
refer to dental unit instructions.
Use of the bur wrench is required to open and close the chuck when installing and removing burs.
Installing and Removing Burs: Tradition (REF 750045 / 750044)
Tradition handpieces accept regular friction grip (FG) burs and short shank (SS) burs.
To use bur wrench: Insert square pin of bur wrench through
the handpiece's head cap until it engages the square socket in
the chuck. Fit the two wrench pins into the matching slots in
the head cap. Turn the wrench counterclockwise to open the
chuck, clockwise to tighten.
To install a bur: Open chuck, insert bur all the way into the
chuck. Tighten chuck with bur wrench.
Make sure bur is secure before operating handpiece.
Always use bur wrench to tighten chuck.
To remove bur: Insert the bur wrench into the back of the handpiece head. Rotate the wrench 1/2 turn
counterclockwise, then remove bur.
Note: The water spray will adequately cover bur and diamond cutting instruments up to .906 inches
(23 mm) long.
See Maximum Recommended Bur Length and Warning Information On Reverse Side.
/ Quiet-air
®
®
REF 730014 / 464004)
®
L
®
REF 730014 / 464004)
Bur Wrench Tradition:
(Wrench REF 750119R)
continued on reverse side

Publicité

loading

Sommaire des Matières pour DENTSPLY Midwest Tradition

  • Page 1 • Midwest turbines should not be rebuilt. Replacing bearings only could result in unexpected Insert handpiece into sterilization bag and load into autoclave. (DENTSPLY recommends using sterilization bags failure and patient injury due to different wear factors of new vs. old components.
  • Page 2 The enclosed business reply card (guaranty registration) must be filled out completely and returned within ten To remove bur: Lift lever (past 90°) to open chuck; then remove bur. When the handpiece is not in use, store (10) days of purchase for this guaranty to apply. DENTSPLY Professional makes no other guaranties, expressed or implied.
  • Page 3 • Ne pas remonter les turbines Midwest. Le fait de ne remplacer que les roulements pourrait Insérer la pièce à main dans le sachet de stérilisation et ensuite dans l’autoclave. (DENTSPLY recommande provoquer une panne et blesser le patient car les facteurs d’usure des anciens composants ne d’utiliser des sachets de stérilisation dont au moins une des parois est en papier).
  • Page 4 La carte commerciale-réponse incluse (enregistrement de garantie) doit être remplie entièrement et retournée dans un délai de dix (10) jours à compter de la date d’achat pour que cette garantie s’applique. DENTSPLY Pour retirer la fraise : soulever le levier (à plus de 90°) pour ouvrir le mandrin ; retirer ensuite la fraise.
  • Page 5 Inserte la pieza de mano en la bolsa de esterilización y colóquela en el autoclave (DENTSPLY recomienda el uso de bolsas de esterilización con al menos una cara de papel). Las piezas de mano se deben procesar en el autoclave componentes nuevos y usados.
  • Page 6 Asegúrese de que la fresa esté bien sujeta antes de utilizar la pieza de mano. adjunta y se debe enviar dentro de los diez (10) días de la compra. DENTSPLY Professional no ofrece ninguna Para retirar la fresa: Levante la palanca (más de 90º) para abrir el portafresas y retire la fresa. Cuando no otra garantía, expresa ni implícita.
  • Page 7 Ausfall und zu Personenverletzungen aufgrund unterschiedlicher Beutel • Autoklav Abnutzungsbedingungen neuer bzw. alter Komponenten kommen. Geben Sie das Handstück in den Sterilisationsbeutel und dann in den Autoklaven. (DENTSPLY empfiehlt die Nebenwirkungen Verwendung von Sterilisationsbeuteln mit zumindest einer Papierseite.) Handstücke sollten im Autoklaven 15 Minuten bei einer Sterilisationstemperatur von 132 ºC behandelt werden.
  • Page 8 Ihr Tradition / Tradition•L / Tradition•PB bzw. Quiet-Air / Quiet-air•L Handstück verfügt über eine Garantie gegen Defekte aufgrund fehlerhafter Materialien oder Bearbeitungsfehlern für die Dauer von sechs (6) Monaten ab Kaufdatum, sofern das Handstück (nach Gutdünken von DENTSPLY Professional) gemäß den Anweisungen betrieben und gewartet und keinem offensichtlichen unsachgemäßen Gebrauch oder einer Montage und Entnahme des Bohrers: Störung ausgesetzt wurde.
  • Page 9 Imbustamento • Autoclave dei componenti vecchi e nuovi. Inserire il manipolo nella busta di sterilizzazione e caricare nell’autoclave. (DENTSPLY raccomanda di utilizzare Effetti indesiderati buste di sterilizzazione con almeno un lato di carta.) I manipoli devono essere autoclavati per un tempo minimo di 15 minuti alla temperatura di sterilizzazione di (132ºC/270ºF).
  • Page 10 Inserzione e rimozione delle frese (a discrezione di DENTSPLY Professional) e non risulti che sia stato soggetto a uso improprio, abuso o incidente. Qualsiasi parte che si ritenga difettosa, imperfetta o allentata deve essere restituita al rivenditore, Inserimento della fresa: Sollevare la levetta —...
  • Page 11 Τοποθετείστε την χειρολαβή μέσα στην σακούλα αποστείρωσης και βάλτε την μέσα στο αυτόκαυστο. διαφοράς των παραγόντων φθοράς των νέων έναντι των παλαιών εξαρτημάτων. (Η DENTSPLY συστήνει να χρησιμοποιείτε σακούλες αποστείρωσης που έχουν τουλάχιστον μια χάρτινη πλευρά.) Οι χειρολαβές θα πρέπει να υποβάλλονται σε αποστείρωση εντός κατάλληλου αυτόκαυστου για τουλάχιστον...
  • Page 12 από την ημερομηνία αγοράς της — με την προϋπόθεση ότι η χειρολαβή έχει χρησιμοποιηθεί και τσοκ. Εισάγετε την άτρακτο του γλυφάνου ολοκληρωτικά μέσα στο τσοκ’ στη συνέχεια κατεβάστε τον Μοχλό συντηρηθεί όπως περιγράφεται μέσα σε αυτές τις οδηγίες (κατά την κρίση της DENTSPLY Professional) Ισχύος για να ασφαλίσει το γλύφανο στη θέση του.
  • Page 13 • Турбины Midwest нельзя восстанавливать. Замена только одних подшипников может привести к Положите наконечник в мешок для стерилизации и поместите его в автоклав. (Компания DENTSPLY непредвиденному отказу системы и травмированию пациента, так как степень износа новых и рекомендует применять мешки для стерилизации хотя бы с одной бумажной стороной.) Обработка...
  • Page 14 с момента их приобретения при условии надлежащей эксплуатации и обслуживания наконечников Установка бора. Поднимите рычаг на угол более 90°, чтобы открыть держатель. Вставьте хвостовик бора в соответствии с данными инструкциями (по усмотрению компании DENTSPLY), а также при полностью в держатель, опустите рычаг, чтобы зафиксировать бор на месте.