10.4 Guidance and manufacturer's declaration

Guidance and manufacturer's declaration - elektromagnetic emissions - for all equipment and systems (see 5.2.2.1 c))
The Ocean E-Vip is intended for use in the electromagnetic environment specified below. The customer or the user of the Ocean E-Vip should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions IEC 61000-3-3
Guidance and manufacturer's declaration - elektromagnetic immunity - for all equipment and systems (see 5.2.2.1 f))
The Ocean E-Vip is intended for use in the electromagnetic environment specified below. The customer or the user of the Ocean E-Vip should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
(IEC 61000-4-2)
Electrical fast transient / burst
(IEC 61000-4-4)
Surge (IEC 61000-4-5)
Voltage dips, short interruptions and voltage
variations on power supply input lines
(IEC 61000-4-11)
Power frequency (50/60 Hz) magnetic field
(IEC 61000-4-8)
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration - elektromagnetic immunity - for all equipment and systems that are not life-supporting (see 5.2.2.2)
The Ocean E-Vip is intended for use in the electromagnetic environment specified below. The customer or user of the Ocean E-Vip should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
Electromagnetic environmant - guidance
Group 1
The Ocean E-Vip uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
Class B
The Ocean E-Vip is suitable for use in all establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Class A
Complies
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input / output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5 % U
(> 95 % dip in U
) for 0.5 cycle
T
T
40 % U
(60 % dip in U
) for 5 cycles
T
T
70 % U
(30 % dip in U
) for 25 cycles
T
T
< 5 % U
(> 95 % dip in U
) for 5 sec
T
T
3 A/m
Compliance level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Ocean E-Vip, including
cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmit-
ter.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input / output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5 % U
(> 95 % dip in U
) for 0.5 cycle
T
T
40 % U
(60 % dip in U
) for 5 cycles
T
T
70 % U
(30 % dip in U
) for 25 cycles
T
T
< 5 % U
(> 95 % dip in U
) for 5 sec
T
T
3 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relati-
ve humidity should be at least 30 %.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Mains power quality should be that of a typical com-
mercial or hospital environment.
If the user of the Ocean E-Vip requires continued
operation during power mains interruptions, it is re-
commended that the Ocean E-Vip be powered from
an uninterruptible power supply or battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical com-
mercial or hospital environment.