COOK Medical DP-M350 Mode D'emploi page 5

Masquer les pouces Voir aussi pour DP-M350:
Table des Matières
Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 62
As per the WEEE directive, Cook Incorporated accepts its responsibility to recover and
recycle our devices.
• The equipment that you bought has required the extraction and use of natural resources for its
production. It may contain hazardous substances that could impact health and the environment.
• In order to avoid the dissemination of those substances into our environment and to diminish the
pressure on the natural resources, we encourage you to use the appropriate take-back systems. Those
systems will reuse or recycle most of the materials of your end-of-life equipment in a sound way.
• The crossed-out wheeled bin symbol invites you to use those systems.
• If you need more information on the collection, reuse, and recycling systems, please contact your local
or regional waste administration.
• You can also contact us for more information on the environmental performances of our products.
For more information on where you can send your waste equipment for recycling, please contact the
appropriate European representative below.
Doppler Monitor and Accessories
COOK IRELAND LTD.
O'Halloran Road
National Technology Park
Limerick, Ireland
Phone: +353 61 334440
Fax: +353 61 334441
Email: froche@cook.ie
MAINTENANCE AND FUNCTIONAL CHECKS
WARNING: The Cook Doppler Blood Flow Monitor System (DP-M350) must only be returned to and repaired
by Cook Incorporated personnel at an authorized Cook Incorporated facility. Should the Cook Doppler
Blood Flow Monitor System (DP-M350) require repair, contact Cook Medical Incorporated or your Cook
Medical Distributor for information on returning the device for repair.
WARNING: If required, the DP-M350 lithium-ion battery pack replacement must be performed only by
trained Cook Incorporated personnel at an authorized Cook Incorporated repair facility, as incorrect
replacement could result in an unacceptable risk to untrained personnel, the monitor or the surrounding
environment.
WARNING: The Cook Doppler Blood Flow Monitor System (DP-M350) contains no user serviceable
components and requires no maintenance or calibration.
WARNING: To isolate equipment from mains remove the battery charger from the AC mains
socket.
TEST MODE (TEST BUTTON)
The Test Button is to be depressed (as stated in the suggested INSTRUCTIONS FOR USE), prior to use of the
unit to verify its proper operation. If, while the Test Button is depressed, no sound /tone is produced and
the monitor is properly powered, replace the monitor as it may require repair at Cook Incorporated. Contact
Cook Medical Incorporated or your Cook Medical distributor for information on returning the device for
repair.
Doppler Blood Flow Monitor Channel/Cable Verifi er
The Channel/Cable Verifi er, DP-MCV01, is provided to allow for real-time bedside testing of the Cook
Doppler Blood Flow Monitor individually or while the monitor is connected to a Cook Extension Cable. Use
is recommended if function problems have not been resolved at the probe site by maneuvering the patient
5
Table des Matières
loading

Table des Matières