Electromagnetic Compatibility - arjo Enterprise 5000X E5X Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

9. Electromagnetic Compatibility

Product has been tested for compliance with current regulatory standards regarding its capacity to
block EMI (electromagnetic interference) from external sources.
Some procedures can help reduce electromagnetic interferences:
• Use only Arjo cables and spare parts to avoid increased emissions or decreased immunity which
can compromise the correct functioning of the equipment.
• Ensure that other devices in patient-monitoring and/or life-support areas comply to accepted
emissions standards.
WARNING
Wireless communications equipment such as wireless computer network devices,
mobile phones, cordless telephones and their base stations, walkie-talkies, etc. Can
affect this equipment and should be kept at least 1.5m away from the equipment.
Intended Environment: Professional Healthcare Facility Environment.
Exceptions: HF Surgical Equipment and the RF Shielded room of an ME SYSTEM for magnetic
resonance imaging.
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
Guidance and manufacturer's declaration – electromagnetic emissions
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Electromagnetic environment - guidance
Group 1
This equipment uses RF energy only for its internal
functions. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Class A
This equipment is suitable for use in all establishments
other than domestic and those directly connected to the
Class A
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Complies
55

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Enterprise 5000xE5x

Table des Matières