Duty Cycle Lockout - arjo Enterprise 5000X E5X Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

To recharge the battery, connect the bed to the electricity supply. Allow at
Recharging the
least eight hours to recharge the battery when it is completely discharged.
backup battery
While the battery is recharging, the ACP battery indicator lights yellow.
The indicator will go out when the battery is fully charged.
WARNING
The battery must only be recharged using the built-in charger. Do not use a separate
charger or power supply.
The backup battery must be ventilated while recharging. Do not cover the battery or
obstruct the area around it.

Duty cycle lockout

Continuous operation of the controls may exceed the duty cycle of
the bed's electrical system, causing the indicators above the buttons
to flash. After 30 seconds, the indicators will light and all functions
are locked.
If this happens, wait for at least 18 minutes then follow the unlocking
procedure described in the section Function lockout.
Location of backup battery
41