Preventive maintenance

This product is subject to wear and tear during use. To ensure that it continues to perform within
its original specification, preventive maintenance procedures should be carried out at the intervals
shown.
WARNING
This list indicates the minimum recommended level of preventive maintenance. More
frequent inspections should be carried out when the product is subjected to heavy use
or aggressive environments, or where required by local regulations.
Failure to carry out these checks, or continuing to use the product if a fault is found,
may compromise the safety of both the patient and carer. Preventive maintenance can
help to prevent accidents.
NOTE
Product cannot be maintained and serviced while in use with the patient.
Actions to be done by carer
Daily
Weekly
Check operation of side rails
Check patient handset and cable
Check ACPand cable
Visually check castors
Visually check power supply cord and mains plug
Check the mattress for damage and fluid ingress
Examine the lifting pole, strap and handle (Optional)
Check for damage to the side rail pads (Optional)
Check for correct fitting of the locking pin on 3 bar and 5 bar removable
side rails
Check operation of manual CPR release handles on both sides of the bed
If the result of any of these tests is unsatisfactory, contact Arjo or an approved service agent.