Chapter 15
MAINTENANCE
15.1 Do not open the enclosure of the device to avoid possible electric shock.
Any maintenance and future upgrades to this device must be carried out by per-
sonnel trained and authorized by our company. The repair should be for our com-
pany's original components only.
15.2 Check the integrality of lead lines periodically. The damage of any lead line may
cause no display of corresponding lead or all leads. Clean the lead lines with water
and soap, and disinfect them with 75% alcohol.
15.3 Please store the electrodes properly. Disinfect them with 75% alcohol and replace
them after long use.
15.4 Use the litmusless cleanser to clean the device. Don't immerge the device in the
cleanser.
15.5 Please pull out the power plug when it is power cut. When not used for a long
period, the device should be placed where it is shady, cool, and dry, powered on
every 3 months.
15.6 Frequent maintenance is necessary for this device. Check it every 6 months and
measure it once a year. If the device has stored/used for more than a year, please
measure it before use again.
15.7 The schematic diagram and list of key parts of this device can be provided only to
the service station or personnel authorized by our company.
Transportation and storage
• Transportation: Please carry out according to the regulation of contract.
• The device after packed should be stored indoor where temperature is -40°C~+55°C,
and relative humidity is less than 95%, without corrosive gas and drafty.