Dräger X-am 5100 Notice D'utilisation page 285

Masquer les pouces Voir aussi pour X-am 5100:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
A
Notes on Approval
A.1
Marking
PO Ex ia I X
0 Ex ia IIC T4/T3 X
Dräger Safety
Type: MQG 0020
23560 Lübeck, Germany
C
0158
I M1 / II 1G
Ex ia I/IIC T4/T3 Ma/Ga
Um=4.6V Im=1.3A
BVS 10 ATEX E 080X
IECEx BVS 10.0053X
ANZEx 11.2003X
Intrinsically safe Ex ia, CSA 11 1800517
.
CSA: Class I, Div. 1, Gr. A,B,C,D TC T4/T3
Class I, Zone 0, A/Ex ia IIC T4/T3 /Ga
-20°C Ta
+50/+40°C: see Battery Pack!
For TC T4/T3: see Battery Pack!
Warning: Read manual for safety precautions.
Avertissement: Lire le manuel avant utilisation.
Do not change or charge batteries in haz loc.
1
Serial No.
1
The year of manufacture is indicated by the third letter in the serial number:
Y = 2007, Z = 2008, A = 2009, B = 2010, C = 2011, D = 2012, E = 2013, etc.
Example: Serial No. ARCH-0054: the third letter is C, which means that the
unit was manufactured in 2011.
Dräger X-am 5100
All manuals and user guides at all-guides.com
US
Battery Pack Type ABT 0100
Temperature Class T4
–20 °C  Ta  +50 °C
if used with alkaline batteries
Duracell Procell MN1500
Temperature Class T3
–20 °C  Ta  +40 °C
if used with NiMH batteries
GP 180AAHC (1800 mAh)
if used with alkali batteries
Varta Powerone 4006
Varta Powerone 4106
Panasonic Powerline LR6
Battery Pack Type HBT 0000 and HBT 0100
Temperature Class T4
–20 °C  Ta  +50 °C
Notes on Approval
285

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Mqg 0020

Table des Matières