Menu functions

3.6.5
Battery main alarm
The alarm is indicated by an intermittent alarm
message:
Flashing special symbol »
« on the right side of
the display:
The battery main alarm cannot be acknowledged or cancelled:
The instrument is automatically switched off again after 10 seconds.
Before the instrument is switched off, the visual, audible and
vibration alarms are activated for a short time.
3.6.6

Instrument alarm

The alarm is indicated by an intermittent alarm
message:
Special symbol »
« displayed on the right side of
the display:
The instrument is not ready for operation.
Commission maintenance personnel or DrägerService to rectify
the error.
4
Menu functions
4.1
Calling the Info Mode
In measuring mode, press the
If any warning or fault messages are present, the corresponding
information or error codes will be displayed (see the Technical
Handbook).
Press the
key successively for the next display. The TWA and
OK
STEL peak values and exposure values will be displayed.
If no key is pressed for 10 seconds, the instrument reverts
automatically to measuring mode.
24
All manuals and user guides at all-guides.com
key for approx. 3 seconds.
OK
4.2
Calling the Info-Off Mode
When the instrument is in a deactivated state, press the
The name of the gas, measuring unit and measuring range
limit value are displayed.
Pressing the
key again exits the Info Off mode (or via timeout).
4.3
Quick Menu
Bump test, see "Performing the bump test" on page 21
Fresh air adjustment, see "Performing a fresh air calibration"
on page 25
Display and deletion of the peak values