Cleaning:
For cleaning disconnect the main plug!
Disassemble the reactor vessel:
Remove the reactor vessel from the LR basic/control base.
Remove the reactor cover from the reactor vessel.
Remove the glass tube from the base as described below.
Use a screwdriver to loosen and remove the four screws on
the two clamps.
Now both clamps can be removed. Then remove the glass
tube with handles from the base.
Disassemble the anchor stirrer:
After removing the glass tube from the vessel base, the an-
chor stirrer together with the reactor vessel bottom of the
socket can be taken apart from the vessel base (see Fig. 23).
O-ring
48
All manuals and user guides at all-guides.com

Maintenance and cleaning

Remove both handles from the glass vessel as following.
Reserve the glass tube with handles. Loosen both screws
with screw driver at the bottom of handles (see Fig. 22).
Then, both handles can be removed from the glass vessel.
Fig. 21
Note: the screws can only be loosened and cannot be re-
moved from the handles.
Please pay attention to the O-ring when you disassemble
the reactor vessel.
Open the screw as described in Fig. 24 with the socket
wrench.
Fig. 23
Fig. 22
Fig. 24