The device can be operated in "Remote" mode via the RS 232
interface or the USB interface connected to a PC and with the
laboratory software Labworldsoft
Note: Please comply with the system requirements togeth-
er with the operating instructions and help section included
with the software.
USB interface
The Universal Serial Bus (USB) is a serial bus for connecting
the device to the PC. Equipped with USB devices can be con-
nected to a PC during operation (hot plugging). Connected
devices and their properties are automatically recognized.
Use the USB interface in conjunction with Labworldsoft
operation in "Remote" mode and also to update the firm-
ware.
USB device drivers
First, download the latest driver for IKA devices with USB
interface from http://www.ika.com/ika/lws/download/
usb-driver.zip and install the driver by running the setup
file. Then connect the IKA device through the USB data
cable to the PC.
The data communication is via a virtual COM port. Con-
figuration, command syntax and commands of the virtual
COM ports are as described in RS 232 interface.
RS 232 interface
Configuration
The functions of the interface connections between
the machine and the automation system are chosen
from the signals specified in EIA standard RS 232 in ac-
cordance with DIN 66 020 Part 1.
For the electrical characteristics of the interface and the
allocation of signal status, standard RS 232 applies in
accordance with DIN 66 259 Part 1.
Transmission procedure: asynchronous character trans-
mission in start-stop mode.
Type of transmission: full duplex.
Character format: character representation in accor-
The following table summarises the (NAMUR) commands understood by the IKA control equipment.
Abbreviations used:
X,y =
numbering parameter (integer number)
m =
variable value, integer
n =
value of variable, floating point number
X = 1
Pt100 thermometer (external temperature sensor)
X = 2
temperature (heating block)
X = 3
safety temperature
X = 4
stirring speed
X = 6
safety stirring speed
All manuals and user guides at all-guides.com
Interfaces and output
.
®
for
®
dance with data format in DIN 66 022 for start-stop
mode. 1 start bit; 7 character bits; 1 parity bit (even);
1 stop bit.
Transmission speed: 9600 bit/s.
Data flow control: none
Access procedure: data transfer from the stirrer ma-
chine to the computer takes place only at the comput-
er's request.
Command syntax and format
The following applies to the command set:
Commands are generally sent from the computer (Mas-
ter) to the stirrer machine (Slave).
The stirrer machine sends only at the computer's re-
quest. Even fault indications cannot be sent spontane-
ously from the stirrer machine to the computer (auto-
mation system).
Commands are transmitted in capital letters.
Commands and parameters including successive pa-
rameters are separated by at least one space (Code:
hex 0x20).
Each individual command (incl. parameters and data)
and each response are terminated with Blank CR LF
(Code: hex 0x20 hex 0x0d hex 0x20 hex 0x0A) and
have a maximum length of 80 characters.
The decimal separator in a number is a dot (Code: hex
0x2E).
The above details correspond as far as possible to the rec-
ommendations of the NAMUR working party (NAMUR rec-
ommendations for the design of electrical plug connections
for analogue and digital signal transmission on individual
items of laboratory control equipment, rev. 1.1).
The NAMUR commands and the additional specific IKA co-
mmands serve only as low level commands for communica-
tion between the machine and the PC. With a suitable ter-
minal or communications programme these commands can
be transmitted directly to the equipment. The IKA software
package, labworldsoft
, provides a convenient tool for con-
®
trolling device and collecting data under MS Windows, and
includes graphical entry features, for motor speed ramps for
example.