Error/Warning
Probe Not Supported
Bootloader X.X.XX.XX error
0 days remaining message
(For LDO and LBOD only)
Meter not configured
Meter will not power on or
powers on intermittently
Unable to access Full
Access Options screen
Unable to access Full or
Operator Access Options
screen
Description
Probe disconnected or
connected improperly
Software not updated to
most current version
Problem with probe
HQd meter does not
support IntelliCAL probe
Software not updated to
most current version.
LDO or LBOD sensor cap
used for 365 days
There are 0 days remaining
in the life of the LDO sensor
cap.
Meter set to incorrect date
and time
Software not updated to
most current version
Software error(s)
Batteries are not installed
correctly
Software not updated to
most current version
Damaged meter
Correct password has not
been entered
Software not updated to
most current version
Solution
Tighten the locking nut on the probe connector.
Disconnect the probe and then connect the probe
again.
To download the most current version of the
software, refer to the applicable product page on
the manufacturer's website.
Connect a different IntelliCAL probe to the meter
to verify if problem is with the meter or the probe.
Contact Technical Support.
To download the most current version of the
software, refer to the applicable product page on
the manufacturer's website.
Replace the LDO or LBOD sensor cap and
®
iButton
.
Replace the LDO sensor cap. Calibration will be
allowed. However, the calibration icon and
question mark will appear on the measurement
screen even if the calibration has passed.
1. Disconnect the probe from the meter.
2. Remove the meter batteries.
3. Install the meter batteries properly. Follow the
polarity makings.
4. Set correct date and time in the meter.
5. Connect the probe and verify that message
has been removed.
To download the most current version of the
software, refer to the applicable product page on
the manufacturer's website.
If the meter starts up correctly, back up the Data
Log and Method files.
To download the most current version of the
software, refer to the applicable product page on
the manufacturer's website.
Examine battery orientation to make sure the
batteries follow the polarity markings. Test again.
Clean the battery terminals, then install new
batteries.
Connect AC power adapter and test again.
To download the most current version of the
software, refer to the applicable product page on
the manufacturer's website.
Contact Technical Support.
Contact Technical Support.
To download the most current version of the
software, refer to the applicable product page on
the manufacturer's website.