Télécharger Imprimer la page

Hach MET ONE 6000 Manuel D'utilisation De Base page 12

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Figure 7 5-pin connector
Table 5 5-pin connector wiring
Pin
Description
Pin
1
4
2
5
3
Common (shield ground)
Figure 8 10-pin connector
Table 6 10-pin connector wiring
Pin RS485 unit RS232 unit Pulse unit
1
RS485 A
2
RS485 B
3
RS485 A
RS232 TX
4
RS485 B
RS232 RX
12 English
Description
Unit main power (9–28 VDC, 1 A
maximum)
Common
Analog unit
Ch 1+
24 VDC external loop power
source
Ch 1-
Channel 1 loop out
Ch 2+
Channel 2 loop out
Ch 2-
Channel 3 loop out
Table 6 10-pin connector wiring (continued)
Pin RS485 unit RS232 unit Pulse unit
5
6
Status +
7
Status -
8
Common (shield ground)
9
Unit main power (9–28 VDC, 1 A maximum)
10
Analog unit
Channel 4 loop out
Common

Publicité

loading

Ce manuel est également adapté pour:

Met one 6003Met one 6005Met one 6013Met one 6015