4.3 Set up scales

Foot screws

5. OPERATION

5.1 Weighing

Start weighing

Weigh babies/infants
ATTENTION!
Incorrect measurement with force shunt
If the housing of the scales is resting on some-
thing, e.g. a towel, the weight measurement will
not be correct.
– When setting up the scales, ensure that they
1. Place the scales on a flat, stable surface.
2. Align the scales by turning the foot screws.
1. Check that there is no load on the scales.
2. Press the Start key.
The display will read
through all elements of the display.
The scales are ready for operation when the display
reads
If the device is being operated with a power supply
unit, the symbol
3. Step onto the scales and try not to move.
4. Read off the measured result.
You can use the 2 in 1 function to weigh babies and
infants. Here the child is held by an adult during weigh-
(2 in 1)
ing. Proceed as follows:
1. Check that there is no load on the scales.
2. Ask the adult to step onto the scales.
3. Read off the measured result.
4. Hold down the arrow key (hold/2 in 1) until NET
appears in the display.
The weight is stored.
0. 0 0
only make contact with the ground via the
foot screws.
se[A
0. 0 0
.
will appear in the display.
appears in the display.
English
and then quickly run
Operation • 45