Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 60– 250 mmHg,
diastolic 40 –199 mmHg,
Pulse 30 –180 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 165 mm x W 107 mm x H 50 mm
Weight
Approx. 408 g (without batteries, with
cuff)
Cuff size
22 to 36 cm
Permissible operating
+5 °C to + 40 °C, 15 – 90 % relative air
conditions
humidity (non-condensing)
Permissible storage
- 25 °C to + 70 °C, ≤ 93 % relative air hu-
and transport condi-
midity, 800 –1050 hPa ambient pressure
tions
Power supply
4 x 1,5 V
Battery life
For approx. 250 measurements, de-
pending on the blood pressure level
and/or pump pressure
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF (cuff), CLASS II (when using a
mains part)
AA batteries
27
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notification
to allow for updates.
• This unit is in line with European Standard EN 60601-1-2
and is subject to particular precautions with regard to elec-
tromagnetic compatibility (EMC). Please note that portable
and mobile HF communication systems may interfere with
this unit. More details can be requested from the stated Cus-
tomer Service address or found at the end of the instructions
for use.
• This device is in line with the EU Medical Devices Directive
93/42/EEC, the "Medizinproduktegesetz" (German Medical
Devices Act) and the standards EN 1060-1 (non-invasive
sphygmomanometers, Part 1: General requirements),
EN 1060-3 (non-invasive sphygmomanometers, Part 3:
Supplementary requirements for electro-mechanical blood
pressure measuring systems) and IEC 80601-2-30 (Medical
electrical equipment – Part 2 – 30: Particular requirements
for the safety and essential performance of automated non-
invasive blood pressure monitors).
• The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful
life. If using the device for commercial medical purposes, it
must be regularly tested for accuracy by appropriate means.
Precise instructions for checking accuracy may be requested
from the service address.