Maintenance

6.1

Inspection of the sensor

6.1.1 Visual inspection

To check your sensor, we recommend the following
procedure:
– The contacts of the connector must be dry.
Moisture, corrosion and dirt in the connector can
lead to false readings.
– Check the cable for buckling, brittle areas or
ruptures.
– Before calibration always examine the membrane
foil optically for signs of damage. The foil must
be intact and clean. Dirty membranes should be
wip ed clean using a soft, moist tissue.
h
Note: An undulated membrane has no influence on
the sensor performance, assuming the membrane
is intact.
– The membrane body must be replaced if the
sensor has too long a response time, the reading
is u nstable or subject to drift, and if the sensor
c annot be calibrated or the membrane shows
sign of mechanical damage.
– Check the cathode area for discoloration, conta-
mination or cracks in the glass. If necessary
rinse with demineralized water and clean with a
clean soft brush or soft paper tissue.
a
Attention! Do not use any cleaning agents contai-
ning alcohol. This could damage the sensor or
lead to fault current.
a
Attention! The glass body is fragile and sensitive
to vibration.
InPro 6850 i / 6900 i / 6950 i
52 206 349
InPro 6850 i / 6900 i / 6950 i O
Sensor 12 / 25 mm
2
© 02 / 17 Mettler-Toledo GmbH
Printed in Switzerland

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Inpro 6900iInpro 6950iInpro 6850i sérieInpro 6900i sérieInpro 6950 sérieInpro série

Table des Matières