Operation; Maintenance: Cleaning The Sensor Head; Disposal - Mettler Toledo InPro 86X0ie Manuel D'instruction

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 32
en
20
Installation on a horizontal line
There is a mark on the sensor head which shows the flow
direction of the media. The sensor needs to be installed in
this direction.
Mark for the installation into a pipe
The sensor must be fitted to the pipe at least 2 meters (6.6 ft)
away from stray light sources (such as sight glasses).
5

Operation

In case of any failure or malfunction, the equipment in which
the sensor is installed, needs to be in a safe condition (de-
pressurized, empty, rinse, vent, etc.).
6
Maintenance: Cleaning the sensor head
The measurement principle provides a large degree of com-
pensation for contamination of the sensor head. However,
the contamination can become so severe, depending on the
operating conditions and media, that the compensation is
no longer adequate and the sensor head must be cleaned.
1. P lace the system or equipment in a safe state (depressu-
rized, empty, rinsed, vented or ventilated, etc.)
2. D isconnect electrical power
3. D ismount the sensor from process
4. C lean the sensor head using a soft, lint-free cloth and
a mild cleaning agent
5. I nstall the sensor
7

Disposal

It is the user's responsibility to dispose of the sensor in
a professional manner. This sensor contains
electronic components that require special
disposal without any hazard for humans, and
with no risk of harm to the environment. Please
observe the applicable local or national reg-
ulations concerning the disposal of "waste electrical and
electronic equipment".
InPro 86X0 i e
30 421 373 B
© 06 / 2 021 METTLER TOLEDO Group, CH
InPro 86X0 i e

Publicité

Table des Matières
loading

Table des Matières