Accuracy Table
Temperature (Measurement value/MAX/MIN/AVG)
Thermocouple type
K
*1: In an environment where the temperature of the instrument is ±1°C and stable, the accuracy is specified.
*2: Instrument can be made to display readings in Fahrenheit (°F) by means of special operation.
18
Electric charge detection
During voltage detection, a continuous buzzer sounds and the red warning backlight lights up.
Range (detection sensitivity)
Hi
Lo
* In contact with the insulated wire that is equivalent to IV2 mm
68
Range
Resolution
-40.0°C to 400.0°C
0.1°C
2
-40°F to 752°F*
0.1°F
Detection voltage range*
40 V AC to 600 V AC
80 V AC to 600 V AC
1
Accuracy*
±0.5% rdg.±3.0°C
±0.5% rdg.±5.4°F
Detection target frequency
50 Hz/60 Hz
50 Hz/60 Hz
2
.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Cm4372Cm4373Cm4374Cm4371-50

Table des Matières