DC current (Measurement value/MAX/MIN/AVG)
(Accuracy guarantee
Range
range)
600.0 A (±1.0 A to ±600.0 A)*
2000 A
(±10 A to ±2000 A)
Auto range movement threshold: 6000 counts or more for upper range, 540 counts or less for lower range.
* For 30.0 A or less, add 0.5 A to the measurement accuracy.
DC current (PEAK MAX/PEAK MIN)
(Accuracy guarantee
Range
range)
600.0 A (±10 A to ±1500 A)
2000 A
(±10 A to ±2300 A)
(±2301 A to ±2840 A)
Measurement accuracy
Resolution
(Values apply regardless of whether the filter is on or off.)
0.1 A
±1.3% rdg.±0.3 A
1 A
±1.3% rdg.±3 A
Resolution
Measurement accuracy
1 A
±1.3% rdg.±7 A
1 A
±1.3% rdg.±7 A
1 A
±6.0% rdg.±7 A
Accuracy Table