Accuracy Table
Frequency (Measurement value/MAX/MIN/AVG)
Range
(Accuracy guarantee range)
9.999 Hz
(1.000 Hz to 9.999 Hz)
99.99 Hz
(1.00 Hz to 99.99 Hz)
999.9 Hz
(1.0 Hz to 999.9 Hz)
Auto range movement threshold: 9999 counts or more for upper range, 900 counts or less for lower range.
13
Continuity check (same for all models)
Continuity on threshold:
Continuity off threshold:
Conditions of guaranteed accuracy:
Response time:
Overload protection:
Overload current:
(Accuracy
Range
guarantee range)
Ω
Ω
600.0
(0.0
to 600.0
64
Ω
Ω
25
±10
(continuous buzzer sound, red warning backlight lights up)
Ω
Ω
245
±10
After zero adjustment has been performed
Detection of open or short for 0.5 ms or more
Lower of 1700 V/1000 V AC or 2×10
30 mA or less at steady state, 1.5 A or less at transient state
Measurement
Resolution
current
Ω
Ω
) 0.1
200 µA±20%
Resolution
Measurement accuracy
0.001 Hz
±0.1% rdg.±0.003 Hz
0.01 Hz
±0.1% rdg.±0.01 Hz
0.1 Hz
±0.1% rdg.±0.1 Hz
7
V • Hz (energized for 1 minute)
Measurement
accuracy
±0.7% rdg.±0.5
Open terminal
voltage
Ω
2.0 V DC or less