Hioki CM4371 Mode D'emploi page 47

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Accuracy specifications and measurement specifications
Guaranteed accuracy period
Guaranteed accuracy period
after adjustment made by Hioki
Guaranteed accuracy for
temperature and humidity
Temperature characteristic
Other conditions
AC measurement method
Conditions of AC accuracy
guarantee
1 year "Accuracy shown in accuracy table"
3 years (reference values) "Accuracy shown in accuracy table ×1.5"
1 year
23°C±5°C (73°F±9°F), 90% RH or less (no condensation)
Within the operating temperature range, add "measurement accuracy ×
0.1/°C" (excluding 23°C±5°C (73°F±9°F)).
When using the L4931 Extension Cable Set, accuracy is guaranteed for
up to two connected cables (totaling 3 m in length).
True RMS measurement
Sine wave input
Specifications