Storage, Disposal - Tecsis PEX10 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

10. Trouble shooting / 11. Storage, disposal
Failure
Signal span dropping off/too small
Signal span drops off
Signal span too small
Signal span erratic
In case of unjustified reclamation we charge the reclamation handling expenses.
*)
Make sure that after the setting the unit is working properly. In case the error continues to exist send in the instrument for reparation (or
replace the unit).
If the problem persists, contact our sales department.
USA, Canada: If the problem continues, contact tecsis or an authorized agent for assistance.
If the pressure sensor must be returned obtain an RMA (return material authorization) number
and shipping instructions from the place of purchase. Be sure to include detailed information
about the problem. Pressure sensors received by tecsis without a valid RMA number will not be
accepted.
Process material certificate (Contamination declaration for returned goods)
Purge / clean dismounted instruments before returning them in order to protect our employees
and the environment from any hazard caused by adherent remaining media.
Service of instruments can only take place safely when a Product Return Form has been
submitted and fully filled-in. This Return Form contains information on all materials with which
the instrument has come into contact, either through installation, test purposes, or cleaning.
You can find the Product Return Form on our internet site (www.tecsis.de).

11. Storage, disposal

When storing or disposing of the pressure sensor, take precautions with
!
regard to remaining media in removed pressure sensors. We recommend
cleaning the pressure sensor properly and carefully. Remaining media in the
Warning
pressure port may be hazardous or toxic!
42
Betriebsanleitung/Operating instructions/Mode d'emploi/Instrucciones de servicio
Betriebsanleitung/Operating instructions/Mode d'emploi/Instrucciones de servicio
Possible cause
Diaphragm is damaged, e.g. through
impact, abrasive/agressive media;
corrosion of diaphragm/pressure
connector; transmission fluid missing.
Seal/Sealing face damaged/conta-
minated, seal mounted incorrectly,
threads crossed.
Mechanical overload through overpres-
sure
Violent fluctuations in the process
media pressure
Procedure
Contact the manufacturer and replace
the instrument
Clean the seal/sealing face, possibly
replace the seal.
Re-calibrate the instrument *)
Damping; consult with manufacturer
GB