Maintenance - Dräger FPS 7000 Notice D'utilisation

Masquer les pouces Voir aussi pour FPS 7000:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

(3) Leak test under water
Immerse the test head including the facepiece with exhalation
valve test plug in the water.
Produce a positive pressure of approx. 10 mbar (approx. 10cm
H
O). Slowly turn the test head under the water surface. Emerging
2
bubbles indicate untight places.
Take the test head with the facepiece out of the water, repair the
facepiece or replace leaking components.
Remove the exhalation valve test plug and mount the exhalation
valve.
Repeat leak test (1).
Checking the opening pressure of the exhalation valve
Check the opening pressure using one of the mentioned test
devices.
The opening pressure must amount to at least 4.2 mbar
O) at a constant flow of 10 l per minute.
(approx. 4.2 cm H
2
If the opening pressure is not reached, turn the valve disk to get
rid of any particles lodged between crater and valve disk.
Repeat the test.
If the opening pressure is still not reached, replace the spring
(figure A2, 17) and repeat the test.
Use Instructions for NIOSH-approved full facepieces

Maintenance

Check all parts for deformation and damage and replace them if
necessary.
After maintenance work and/or replacement of components again
check the leak tightness.
Replacing the visor (figure F)
Remove the screw from the flap and take off the flap.
Unhook the cap from the visor frame (Figure F1-1) on both sides.
Use your thumb to press against the bottom part of the cap, then
remove the cap from the fastening noses (Figure F1-2).
Pull the cap upwards and away from the facepiece.
Remove the screws of the visor frame (Figure F2).
Spread the frame apart, tilt it towards the front and pull apart
sideways.
Detach the visor from the rubber frame.
Moisten either the facepiece body or the visor frame to facilitate
the assembly.
Insert the new visor centered by aligning it to the center marking
on the visor and form-fitting in the facepiece body.
Apply pressure onto the visor frame until it snaps into place, then
tighten with new self-locking screws until there is no more gap.
Put the cap onto the connection piece, lock it into place on the
fastening noses and hook it into the visor frame.
Close the flap and fasten the screw.
Replacing the nose cup (figure G)
Take hold of the nose cup at the tip of the nose, then pull it off the
pin and out of the speech diaphragm groove.
Put the new nose cup onto the pin to center it.
Press the ring of the nose cup into the speech diaphragm groove
- at first at the bottom and then all around the speech diaphragm.
11