Hach WDMP sc Manuel De L'utilisateur page 13

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

For installation with conduit:
Note: To supply AC power to the controller with conduit, refer to the controller user manual.
• Install a local disconnect switch or circuit breaker with sufficient current capacity for the panel and
that breaks all mains powered (hot) conductors.
breaker. Put a label on the disconnect or circuit breaker that identifies it as the main disconnect
device for the instrument. Make sure the local disconnect device is within easy reach of the user
and easy to operate.
• Connect the conduit through a conduit hub that holds the conduit securely and seals the enclosure
when tightened.
• Make sure that the wire insulation has a minimum rating of 300 V, 80 ºC.
• Make sure that the ground wire is 18–14 AWG and connects uninterrupted to the ground stud. Put
an earth ground symbol label on the ground stud.
Mains circuit
115–120 VAC
230 VAC
1. Remove the controller and brackets. Refer to the illustrated steps that follow.
Table 3
Table 3 Circuit breaker requirements
shows the requirements for the circuit
Amps
≤ 20 A
≤ 10 A
English 13

Publicité

Table des Matières
loading

Table des Matières