1.2 Combination possibilities
This prosthetic component is compatible with Ottobock's system of modular
connectors. Functionality with components of other manufacturers that have
compatible modular connectors has not been tested.
2 Intended use
2.1 Indications for use
The product is intended exclusively for lower limb exoprosthetic fittings.
2.2 Area of application
Our components perform optimally when paired with appropriate compon­
ents based upon weight and mobility grades identifiable by our MOBIS clas­
sification information and which have appropriate modular connectors.
The product is recommended for mobility grade 1 (indoor walk­
er) and mobility grade 2 (restricted outdoor walker).
kg
The maximum approved body weight is specified in the technical data
(see page 19).
2.3 Environmental conditions
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative humidity 20 % to
90 %, no mechanical vibrations or impacts
Allowable environmental conditions
Temperature range: –10 °C to +45 °C (14 °F to 113 °F)
Chemicals/liquids: fresh water, salt water, perspiration, urine, soapsuds, chlorine
water
Moisture: Submersion: max. 1 h in 3 m depth, relative humidity: no restrictions
Solids: Dust, occasional contact with sand
Clean the product after contact with humidity/chemicals/solids, in order to
avoid increased wear and damage (see page 18).
Unallowable environmental conditions
Solids: Highly hygroscopic particles (e.g. talcum), continuous contact with sand
Chemicals/liquids: Acids, continuous use in liquid media