Télécharger Imprimer la page

Roche Accu-Chek Aviva Nano Manuel De L'utilisateur page 32

Masquer les pouces Voir aussi pour Accu-Chek Aviva Nano:

Publicité

Les langues disponibles

Les langues disponibles

4
Control Tests
Understanding Out-of-Range Control Results
WARNING
The control ranges are printed on the test strip container label. If the control result is out of range, check
this list to help solve the problem.
Troubleshooting Checks
1. Were the test strips or control solutions
expired?
2. Did you wipe the tip of the control solution
bottle before use?
3. Were the caps on the test strip container
and the control solution bottle always
closed tightly?
4. Was the test strip used immediately after it
was removed from the test strip container?
5. Were the test strips and control solutions
stored in a cool, dry place?
6. Did you follow the directions?
7. Did you choose the correct control solution
level, either 1 or 2, when you performed the
control test?
8. Are you still unsure of the problem?
30
Action
Discard the test strips or control solution if either is
past the use by date. If the control solution was
opened more than 3 months ago, discard it. Repeat
the control test with an unexpired test strip and an
unexpired control solution.
Wipe the tip of the bottle with a tissue. Repeat the
control test with a new test strip and a fresh drop of
control solution.
Replace the test strips or control solution if you think
either was uncapped for some time. Repeat the
control test.
Repeat the control test with a new test strip and a
fresh drop of control solution.
Repeat the control test with a properly stored test
strip or control solution.
Read Chapter 4, Control Tests, and repeat the control
test.
If you chose the wrong control solution level, you can
still compare the control result to the range printed
on the test strip container.
Contact Accu-Chek Customer Care at
1-800-363-7949.

Publicité

loading

Ce manuel est également adapté pour:

Accu-chek aviva