Télécharger Imprimer la page

Ottobock 1C66 Triton smart ankle Instructions D'utilisation page 29

Masquer les pouces Voir aussi pour 1C66 Triton smart ankle:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Activation
1) Extend the prosthetic foot forward so that the angle between the sole of the foot and the floor
is at least 30°.
2) Place a load on the forefoot area in this position for at least 2 s (e.g., by pushing against a
wall).
→ 3 short beeps occur. The ankle joint of the prosthetic foot is locked.
Deactivation
► Place the prosthetic foot flat on the floor and place a load on it for at least 1 s.
→ 1 short beep occurs. The ankle joint is unlocked.
6.3 Warning signals
The prosthetic foot emits warning signals to indicate problems. There are 3 categories of warning
signals.
Warning signal without functional limitation
The prosthetic foot works with no limitations. However, user action is required to avoid errors.
Possible errors
A detailed error message and action instructions are displayed in the Galileo™ app.
Warning signal with functional limitation
The functionality of the prosthetic foot is restricted. The prosthetic foot no longer adapts to dif­
ferent situations.
Possible errors
A detailed error message and action instructions are displayed in the Galileo™ app.
Warning signal with deactivation of all functions
All functions of the prosthetic foot are deactivated. The functionality can only be restored
through intervention by the user or orthopaedic technician.
Possible errors
The Bluetooth connection is severed.
7 Cleaning and Care
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
3) Allow to air dry in order to remove residual moisture.
Charge level low, maintenance required, fault in the hydraulics
Energy saving mode (LED flashing yellow): The battery
charge level has dropped so low that the prosthetic foot has
activated the energy saving mode.
Charge the battery to restore full functionality (LED flashes
green rapidly for 1 second).
Temperature warning, hydraulics: The temperature of the
hydraulics is too high or too low. As soon as the temperature is
within the allowable range again, full functionality is restored.
Battery fully discharged: The prosthetic foot turns off. The
current position is maintained.
Fully charge the battery and press the main switch to turn the
prosthetic foot back on.
System error (LED alternately flashes green and yellow):
The prosthetic foot no longer adapts automatically. By pressing
the selection key, the foot position can be adjusted manually.
Contact the manufacturer to discuss the further procedure.
3 beeps every 3 minutes
LED flashes yellow rapidly
5 beeps every 3 minutes
10 beeps
Ottobock | 29

Publicité

loading